Model Number 6947M55 |
Device Problem
Decreased Sensitivity (2534)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: 479878 lead implanted: (b)(6) 2020; 407645 lead implanted: (b)(6) 2014 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient presented to the emergency department and reported that their heart rate was in the forties while using their oximeter.The patient also reported a fast heart rate on that same day.The patient was on telemetry at the emergency department and they did not see any low or fast rates.The physician noted it appeared the patient may have had premature ventricular contractions (pvc)/premature atrial contraction.The right ventricular (rv) lead exhibited chronically low r waves.The cardiac resynchronization therapy defibrillator (crt-d) and rv lead remain in use.No further patient complications have been reported as a result of this event.
|
|
Event Description
|
It was further reported that there were no product performance issues with either the device or rv lead.No tachycardia or bradycardia was confirmed in the emergency department.The patient has frequent pvcs an is likely why pulse oximeter showed low hart rate.No reprogramming was necessary.The patient's medication was adjusted due to the hypotension.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|