Model Number 8100
Device Problems Contamination (1120); Peeled/Delaminated (1454); Insufficient Flow or Under Infusion (2182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted once the failure investigation has been completed.Per 803.52(f)(11)(iii) the information provided represents all of the known information at this time.The complainant or reporter was unable or unwilling to provide any further patient, product, or procedural details to the manufacturer.
 
Event Description
It was reported that there was an under infusion of an unspecified medication.The pump recorded 136.2ml while the actual fluid volume was 128.8ml.An additional 7.4ml of flow was recorded to empty the infusion bag.There was patient involvement but no harm.
 
Manufacturer Narrative
Omit : b21 - type of investigation not yet determined, c21 - results pending completion of investigation, d16 - conclusion not yet available.Additional information : unique identifier (udi) #, device eval by manufacturer?, reason code for no evaluation, if other specify, imdrf annex a, b, c, d, g codes , remedial action required, remedial action # and manufacturer narrative.A device history record review is performed on each device reported in a mdr reportable event along with other methods of investigation as coded in section h6 of this mdr report.H3 other text : not applicable.Device evaluated by bd.
 
Event Description
It was reported that there was an under infusion of an unspecified medication.The pump recorded 136.2ml while the actual fluid volume was 128.8ml.An additional 7.4ml of flow was recorded to empty the infusion bag.There was patient involvement but no harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SYSTEM
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
brett wilko
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key18872365
MDR Text Key337294669
Report Number2016493-2024-19066
Device Sequence Number1
Product Code FRN
UDI-Device Identifier10885403810015
UDI-Public(01)10885403810015
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133532
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation 505
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8100
Device Catalogue Number8100 ALARIS LVP MODULE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Date Manufacturer Received02/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-2718-2020
Patient Sequence Number1
Treatment
8015; 8100
-
-