Catalog Number 8065753041 |
Device Problem
Reflux within Device (1522)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that an ophthalmic console portional reflux was not working.Details of procedure and patient harm was not provided.Additional information has been requested and none received till date.Additional information received reporting that there was no patient harm.
|
|
Manufacturer Narrative
|
The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The system was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|