GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 620; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Inaccurate Delivery (2339)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The link-25 system was re-installed and calibrated to resolve the issue.Block a: no report of patient involvement.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
 
Event Description
It was reported that there was a malfunction that could cause a hypoxic mixture in the patient circuit.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 620
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH   214028
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key18874740
MDR Text Key337327738
Report Number9710602-2024-00336
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier00840682103985
UDI-Public01008406821039854
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-