BAXTER HEALTHCARE CORPORATION SOLUTION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number EMC9608
Device Problems Material Puncture/Hole (1504); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2023
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of a solution administration set had a hole resulting in air bubbles.This was observed during priming.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information: h3, h4, h6, h10.H10: the actual device was not available; however, three photographs of the sample and two retained samples were provided for evaluation.The returned photographs were reviewed, and although liquid droplets were present on the outside of the tube, it was not possible to identify the position of the cut/hole from the photographs provided.The reported condition was verified for the photographs of the sample.The cause of the reported condition could not be determined.Visual inspection on the two retention samples did not identify any abnormalities that could have contributed to the reported condition.Functional tests (push-pull test, underwater leak test, air passage test) were performed, and no disconnections were identified, no leaks were identified, and no blockages were found.The retention samples underwent traction testing, and it was confirmed that both samples withstood the minimum required force.The reported condition was not verified on the retained samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLUTION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18875823
MDR Text Key337337419
Report Number1416980-2024-01044
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMC9608
Device Lot Number23C15V814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-