It was reported that a few days after implant of the cardiac resynchronization therapy defibrillator (crt-d) system, the left ventricular (lv) pace impedance was less than 200 ohms and there was loss of capture in certain programming configurations.The impedance measurement the day prior was 440 ohms.Some noise was observed when measuring impedance and the lv threshold was highly variable.Engineering reviewed the device data and determined there was an issue with the lv pacing channel of the crt-d device and immediate device replacement was recommended.The patient had experienced syncope and was re-admitted to the hospital.It was noted that the patient had upcoming chemotherapy appointments and there was also a large hematoma over the device implant site.The device was subsequently replaced and is expected to be returned for analysis.No additional adverse patient effects were reported.It was additionally reported that the pocket hematoma was drained during the device replacement procedure.The lv lead was tested in both bipolar and unipolar configurations and capture was confirmed, and impedances were stable.The lv lead remains in service.
|