Catalog Number AB2000
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Manufacturer Narrative
H.6 adverse event problem component code 4756: per the instructions for use, the aquabeam motorpack, a re-usable component of the aquabeam robotic system, provides power to the aquabeam handpiece by means of dc motors.Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during aquabltion procedural setup, the aquabeam robotic system generated an "e23 - motorpack error" and could not be cleared despite multiple troubleshooting steps.A second aquabeam motorpack was opened but the issue persised.A second aquabeam handpiece was opened and procedure was completed successfully.The reported event caused a surgical delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
H.11 additional manufacturer narrative: the aquabeam handpiece was returned for investigation.The aquabeam handpiece was tested with the associated aquabeam motorpack triggering e22 and e23 error messages and could be cleared.Additionally, the handpiece lower shell and pcb were swapped with a known good one and the handpiece was tested with the related motorpack.Three docking cycles and an entire simulated aquablation session were performed successfully without triggering any errors, indicating a faulty pcb.Root cause is pcb failure and root cause source is undeterminable.A review of the device history record (dhr) for the aquabeam robotic system/serial number (b)(6) and aquabeam handpiece/ lot number 23c04637 was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the device met all design and manufacturing specifications when released for distribution.The current user manual sj-um0101-00 rev.B, um, aquabeam robotic system user manual, us, baytech was reviewed.The aquabeam robotic system user manual, um0101 rev.F, states the following: table 5: system detected errors and faults e23 - motorpack error release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech dr
san jose, CA 95134
6502327291
MDR Report Key18881619
MDR Text Key337531329
Report Number3012977056-2024-00071
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D202203306
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/30/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-