STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN TRIATHLON A29 ASYMMETRIC METAL BACKED PATELLA; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number UNK_JR |
Device Problems
Degraded (1153); Unstable (1667)
|
Patient Problems
Pain (1994); Joint Laxity (4526)
|
Event Date 02/19/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Reported event: an event regarding instability and wear involving an unknown triathlon patella was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
|
Event Description
|
It was reported that the patient's right knee was revised due to flexion instability and pain.Intra-operatively, the following was noted: the 'lip' of the liner at the posterior medial aspect showed wear.The poly portion of the metal-backed patella showed abnormal wear and was reported that it was possibly clicking against the femoral component.The patient's knee construct was revised to a ts knee with stems and augments.
|
|
Search Alerts/Recalls
|
|
|