Model Number 40025 |
Device Problems
Perivalvular Leak (1457); Incomplete Coaptation (2507); Central Regurgitation (4068)
|
Patient Problems
Calcium Deposits/Calcification (1758); Valvular Insufficiency/ Regurgitation (4449); Aortic Valve Insufficiency/ Regurgitation (4450)
|
Event Date 01/23/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that 26 days post implant of this 25mm aortic bioprosthetic valve, it was explanted and replaced with a different manufacturers product.The reason for the replacement was reported as moderate to severe aortic insufficiency (ai) and paravalvular leak of the prosthetic.It was also reported that an echocardiogram (echo) discovered the non-coronary and left coronary leaflets were "hypomobile and with flattering" and no physiologic movement.No additional adverse patient effects were reported.
|
|
Event Description
|
Medtronic received additional information which stated that the physician believed that an excessive amount of calcium on the aortic root probably did not allow for the correct positioning of the valve on the supra annular plane.It was stated that the valve was explanted due to questionable echocardiography.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|