• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C45101340-NLJ
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the anesthesia bag could not be used because a leak occurred when it was connected to the anesthesia machine.There was no patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
Investigation summary: three photos were included in attachments for evaluation; pinhole was detected in the anesthesia bag.One sample was also received, without its original packaging.Per visual inspection a pinhole was detected in the bag, confirming the customer's complaint.A capa was opened to address the root cause of the investigation.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18886198
MDR Text Key337749245
Report Number9617604-2024-00222
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberC45101340-NLJ
Device Lot Number4427493
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/11/2024
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-