Model Number D97130F5
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226); Cardiac Perforation (2513)
Event Type  Injury  
Event Description
It was reported that right ventricular perforation occurred when the catheter was inserted during a tavi procedure.The customer noticed decreased blood pressure and fluoroscopy revealed cardiac tamponade.Pericardiocentesis was performed.The customer was suspicious that the adverse events were attributable to the catheter because the catheter tip felt stiffer than usual.The catheter was discarded.Patient demographic information, outcome, and status was requested but is unknown, further information could not be obtained as reporter declined to be interviewed.
 
Manufacturer Narrative
The device was not returned for evaluation.Without the return of the device it is not possible to assess any manufacturing issues with the device.The lot number was not provided thus a device history record was not reviewed.The instructions for use , under complications perforation of the right ventricle states as follows, cases of myocardial perforation associated with the use of temporary transvenous pacing catheters have been reported.Careful repositioning and withdrawal of the catheter under ecg and fluoroscopic control is recommended.The instructions for use has been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions or conditions for the successful use of the device.Complaint histories for all reported events are reviewed through trending on a monthly basis and continue to be monitored for any unfavorable trends and documented as part of this monthly review.No corrective or preventative actions are required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco 00610
*  00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco 00610
*   00610
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key18886256
MDR Text Key337444669
Report Number2015691-2024-01953
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K822723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberD97130F5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-