Catalog Number 367364
Device Problem Difficult or Delayed Activation (2577)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2023
Event Type  malfunction  
Event Description
It was reported that while using bd vacutainer® ultratouch¿ push button blood collection set, the needle didn't retract after activation of the button.No patient impact reported.
 
Manufacturer Narrative
D2a.Common device name: blood specimen collection device; intravascular administration set d2b.Medical device type: jka h.6.Investigation summary: material #: 367364 lot/batch #: 3226750 bd received one sample for investigation.The sample was evaluated by functional testing and the indicated failure mode for does not retract with the incident lot was not observed.Additionally, thirty (30) retention samples from bd inventory were evaluated by functional testing and no issues were observed as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode does not retract.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET
Type of Device
SEE H.10
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18886570
MDR Text Key337533982
Report Number1024879-2024-00219
Device Sequence Number1
Product Code FPA
UDI-Device Identifier50382903673648
UDI-Public(01)50382903673648
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K212724
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number367364
Device Lot Number3226750
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-