• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE CORE SMART CABLE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE CORE SMART CABLE; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0600-0783
Device Problems Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/05/2024
Event Type  malfunction  
Manufacturer Narrative
The customer declined the option to have their glidescope core smart cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Review of complaint history for the reported smart cable serial number (b)(6) did not identify any previous complaints reported to verathon.Trending analysis for the glidescope core smart cables does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
A customer reported that during a patient procedure, using a glidescope core smart cable, the image was intermittent.Furthermore, it was reported that there was an unspecified amount of delay intubating the patient.The procedure was completed using a different glidescope core smart cable which was made available in an unspecified amount of time.No harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9,
CA  V5C 5A9,
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9,
CA   V5C 5A9,
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key18886579
MDR Text Key337644897
Report Number9615393-2024-00050
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0600-0783
Device Catalogue Number0800-0602
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/15/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-