Model Number 1871 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the catheter tip was crumbling.A 6f guidezilla ii guide extension catheter was selected for use.Upon unpacking, the tip of the catheter was found damaged and crumbling.The procedure was then completed with another of the same device and there were no patient complications reported.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.A visual inspection revealed numerous flat shaft segments of which included the tip of the device.A microscope inspection revealed no further damage or irregularity.There were no fluids detected to the device.
|
|
Event Description
|
It was reported that the catheter tip was crumbling.A 6f guidezilla ii guide extension catheter was selected for use.Upon unpacking, the tip of the catheter was found damaged and crumbling.The procedure was then completed with another of the same device and there were no patient complications reported.
|
|