It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited right atrial (ra) lead noise with oversensing.Review of the presenting electrogram (egm) showed two signals, one small signal followed by a larger signal that looked like a true p-wave.Review of ra lead trends showed low amplitude p-waves as small as 0.3 mv with other measurements around 2.5-3 mv.Technical services (ts) discussed troubleshooting options, programming considerations, and possible causes.It was noted that the patient had a history of atrial fibrillation (af) and x-rays showed that the ra lead was low, on the right side, and near the valve.There were no known ablations or other cardiac procedures outside of the device implant.The patient was scheduled for a office visit on the day of report and the plan was to address these observations via reprogramming.This crt-p system remains in service.No adverse patient effects were reported.
|