• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G124
Device Problems Over-Sensing (1438); Failure to Read Input Signal (1581)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Event Description
It was reported that the physician contacted boston scientific rhythmcare for recommendations regarding reprogramming this cardiac resynchronization therapy defibrillators (crt-d).Upon data review, it was discussed that a recently recorded ventricular tachycardia (vt-1) episode was inappropriately declared, and was most likely an atrial arrhythmia in origin.However, this episode occurred beneath the programmed therapy zone, and thus no inappropriate therapy was delivered.Rhythmcare provided potential programming options, such as adjusting the rhythmid feature, or raising the vt therapy zone, if clinically appropriate.Additionally, the physician could elect not to alter the programmed settings, as this rate fell below the vt therapy zone.At this time, the crt-d remains implanted and in-service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18891420
MDR Text Key337527833
Report Number2124215-2024-15093
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG124
Device Catalogue NumberG124
Device Lot Number511990
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
-
-