During the pulmonary vein isolation (pvi) procedure, the patient experienced a stroke.The cavo-tricuspid isthmus, pulmonary vein isolation, roof linear line, vein of marshall were performed in sequence, and finally, peri-mitral isthmus (pmi) ablation was performed.While attempting the pmi ablation near the annulus, a high temperature and high impedance alarm was noted (above 180, ablation #46, during ablations #46-50 impedance of 220-240).The cable was exchanged, and another grounding patch was applied.Upon checking the isolation of the pvi, an alarm indicating high impedance was noted, which was attributed to the presence of a pouch near the vein (anterior roof side).The alarm continued so the catheter was removed from the patient and char was noted at the tip of the catheter.Immediately after the char was confirmed that patient was transferred to the intensive care unit and an mri and ct scan were performed.The day after the procedure, the patient displayed a "drowsy condition" due to stroke (multi-infarction) and the patient's condition has remained the same.
|
Four images were submitted for evaluation, no product was received.The returned images appeared to show char on the catheter tip.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the char on the catheter tip could not be conclusively determined.
|