Model Number 459888 |
Device Problems
Fracture (1260); Device Contamination with Body Fluid (2317); Activation, Positioning or Separation Problem (2906); Device Dislodged or Dislocated (2923); Positioning Problem (3009)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/19/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that placement difficulty was experienced with the attempted left ventricular (lv) lead which dislodged upon removing the guidewire.An attempt was made to advance the guidewire again but was unsuccessful.The lead was removed.Blood was observed on the lead and the lead was reported to have possibly fractured.A different lead was used to complete the implant.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the full lead was returned and analyzed.Analysis indicated the distal tip of the lead was damaged.The tip seal on the distal electrode of the lead showed evidence of blood ingression.Visual analysis of the lead indicated damage at implant.The analyst noted visual analysis showed the tip seal dislocated inside the lead tip nose.Therefore, back loading of guidewire could not be performed.By front loading, the guidewire passed the is-4 pin and the lead coil lumen but won¿t passed through the distal end at the location of the tip seal damaged.This indicated no conductor distorted and no blood obstruction in coil lumen.There was no insulation breach was observed.By design, left heart leads are manufactured with a septum in the tip that will sometimes allow blood ingress.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|