Catalog Number CDS0706-XTW
Device Problem Incomplete Coaptation (2507)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  Injury  
Event Description
It was reported that a mitraclip procedure was performed on (b)(6) 2024, to treat degenerative mitral regurgitation (mr) grade 4+.A xtw lot: 30915a1079 was implanted laterally reducing mr to grade 2+.A second xt mitraclip (lot:30502r1062) was then placed laterally.During capturing attempts, the xtw clip was observed to be detached from the anterior leaflet (single leaflet device attachment (slda)).No contact between the two clips was observed.The xt was then implanted to stabilize the slda, along with a nt mitraclip (lot: 30831r1106).Fourth clip ntw (30425r1013) was then implanted to further reduce mr.The procedure ended with mr reduced to grade 1.In the physician's opinion the slda was related to the patient's redundant posterior leaflet.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the reported slda was due to challenging patient anatomy.The reported unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL NORTHPOINT REG 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18894442
MDR Text Key337550583
Report Number2135147-2024-01101
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public05415067037381
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0706-XTW
Device Lot Number30915A1079
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-