• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG Back to Search Results
Model Number 6662
Device Problem Wireless Communication Problem (3283)
Patient Problems Failure of Implant (1924); Movement Disorder (4412)
Event Date 02/19/2024
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.E1: reporter phone number (b)(6).
 
Event Description
It was reported that the patient¿s ipg does not communicate with external devices following unrelated surgery.Reportedly, the ipg was not set into surgery mode prior to the procedure.Troubleshooting was unable to resolve the issue.As such, surgical intervention may take place at a later date to address the issue.
 
Event Description
Date of device implant received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR
Type of Device
DBS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key18897188
MDR Text Key337587328
Report Number1627487-2024-07524
Device Sequence Number1
Product Code MHY
UDI-Device Identifier05415067020260
UDI-Public05415067020260
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/08/2023
Device Model Number6662
Device Lot Number8050507
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-