AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DCB00
Device Problems Break (1069); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Manufacturer Narrative
Section a2, a4 and a5: unknown/ asked but not available.Section d6a: if implanted, give date: unknown, information not provided.Section d6b: if explanted, give date: unknown, information not provided.Section h3: other 81: the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that there was issue in the advancement of the preloaded intraocular lens(iol) due to bent haptic.The issue was observed during handling and before patient contact.No other information is available.
 
Manufacturer Narrative
Corrected data: in review, it was noticed that the "initial reporter email address" was inadvertently not entered in the initial or follow up mdr report #1; therefore, the information has been captured in this supplemental mdr report as indicated below: section e1: initial reporter email address: (b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes section d9: returned to manufacturer on: 03/18/2024 section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection of the complaint simplicity reveal that the lens was stuck in the tip of the cartridge.The cartridge tip was also observed to be damaged.Ophthalmic viscoelastic device (ovd) was not observed to be distributed evenly throughout the cartridge indicating that an insufficient amount of ovd may have contributed to the complaint issues reported.The lens module was opened and no ovd or damage was observed inside.The handpiece was disassembled and no issues that could cause or contribute to the complaint issues were observed.The lens could not be removed from the cartridge.The complaint issue "haptic damaged" and "device advancement issue" were not identified during product evaluation.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: based on the investigation, no malfunction or product deficiency was identified.Corrected data: in review, it was noticed that in the section "g2 " of the initial mdr report, the box (company representative) was inadvertently checked in error which should not have been selected.Also, in review, it was noticed that the following fields were inadvertently entered incorrectly in the initial mdr report.This supplemental mdr is to correct the information.The following fields were updated accordingly: section a2, a4 and a5: not applicable as there was no patient contact.Section d6a: if implanted, give date: not applicable, the lens was not implanted as there was no patient contact.Section d6b: if explanted, give date: not applicable, the lens was not implanted as there was no patient contact.Therefore, not explanted.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key18897204
MDR Text Key337636769
Report Number3012236936-2024-00577
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474636163
UDI-Public(01)05050474636163(17)260312
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDCB00
Device Catalogue NumberDCB0000240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-