ABBOTT QUADRA ASSURA MP ICD; NO MATCH Back to Search Results
Model Number CD3371-40QC
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Under-Sensing (1661)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2024
Event Type  Injury  
Event Description
It was reported that inappropriate high voltage shocks and antitachycardia pacing (atp) for supraventricular tachycardia was observed on the device.Abbott technical support was reviewed device session records and confirmed the event was due to functional undersensing of p-waves.The physician did not allege a malfunction against the device and alleged the device performed as expected based upon it's programmed settings.No intervention has been performed at this time to resolve the event.The patient was in stable condition and will continue to be monitored.
 
Event Description
Additional information received indicated the device was reprogrammed to resolve the event.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18897478
MDR Text Key337590017
Report Number2017865-2024-34900
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberCD3371-40QC
Device Lot NumberP000090204
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-