Model Number CD3357-40Q
Device Problems High impedance (1291); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2024
Event Type  Injury  
Event Description
It was reported that the patient presented for remote follow-up via merlin.Net.A review of the transmission found the right atrial pacing lead impedance above the upper limit.The patient was scheduled for revision.During the procedure the lead was tested via pacing system analyzer and all lead measurements were found to be within normal limits.The physician then determined that the source of the issue was the header of the implantable cardioverter defibrillator.The device was explanted and replaced with acceptable impedance measurements.The patient was discharged.
 
Manufacturer Narrative
The reported event high pacing impedance and header anomaly could not be confirmed.Visual inspection of the septum, setscrew and contact spring did not reveal any anomaly that could contribute to the reported event.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18897965
MDR Text Key337594035
Report Number2017865-2024-34902
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508216
UDI-Public05414734508216
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCD3357-40Q
Device Lot NumberP000188952
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA; LEFT VENTRICULAR LEAD; RIGHT ATRIAL LEAD
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient SexMale
-
-