Model Number G447
Device Problems Failure to Read Input Signal (1581); Low impedance (2285); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  Injury  
Event Description
It was reported that the recently implanted right ventricular (rv) lead exhibited intrinsic amplitude out of range.The rv pacing impedance also has decreased from 530 ohms to 366 ohms and the rv threshold has increased from 0.7v (volts) to 2.1v.A lead assessment with a programmer was recommended.The battery longevity is normal and the other lead measurements are stable.The rv lead and this recently implanted cardiac resynchronization therapy defibrillator (crt-d) remain in service.No adverse patient effects were reported.Additional information provided advised a chest x-ray appeared to show the rv lead had fallen from the septum.The patient will be admitted to the hospital and a lead reposition will be scheduled.At this time, the lead and device remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18899761
MDR Text Key337618574
Report Number2124215-2024-15390
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number315631
Was Device Available for Evaluation? No
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Hospitalization;
Patient Age86 YR
-
-