Model Number HAMILTON-T1
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2022
Event Type  malfunction  
Event Description
Hey (b)(4), so to give more context on the event they experienced: ventilator was plugged in.The user had attached a set to the ventilator prior to turning it on.The two battery leds were illuminated before attempting to power on the ventilator.When user engaged the power button, the two battery leds went out but the screen did not turn on.Instead, a constant audible alarm was triggered and continued to sound until the user held down the power button for roughly 10 seconds.They did try using the silence alarm button but was unsuccessful.After setting aside the ventilator for roughly 2 hours, the user attempted to turn on the device and was successful.Ventilator was working as expected.I have uploaded the service logs, event logs, and error logs.Below are the answers to the questions regarding the ventilator.
 
Manufacturer Narrative
The complaint has been reopened and reviewed according to fda form 483 inspectional observation ems #2, eobs2 from the fda inspection conducted between july 17 to july 21, 2022 at the ems and bonaduz sites.A detailed investigation was performed by an expert from the technical service: this issue is deemed a reportable event since the malfunction may lead to a delay of the therapy as the device has to be re-started.The root cause could not be identified since the startup issue is not reproducible.Within this investigation it has been confirmed that the device failed to meet its specifications at the time of the event.There was no patient or user harm.The allegation in this complaint was confirmed to be a complaint.No further investigation or correction will be performed except those mentioned above.In future hamilton medical ag will report an event similar to this issue as it will be deemed a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMILTON MEDICAL AG
Type of Device
HAMILTON-T1
Manufacturer (Section D)
HAMILTON MEDICAL AG
via crusch 8
bonaduz, 7402
SZ  7402
Manufacturer Contact
andre sonnenberg
via crusch 8
bonaduz, graubunden 7402
SZ   7402
MDR Report Key18900845
MDR Text Key337693507
Report Number3001421318-2023-42433
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00730002856789
UDI-Public0730002856789
Combination Product (y/n)N
PMA/PMN Number
K181216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHAMILTON-T1
Device Catalogue Number161006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/08/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-