• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPATZ FGIA INC. SPATZ3 ADJUSTABLE BALLOON SYSTEM; GASTRIC BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPATZ FGIA INC. SPATZ3 ADJUSTABLE BALLOON SYSTEM; GASTRIC BALLOON Back to Search Results
Model Number A-SP3-03K
Device Problems Patient-Device Incompatibility (2682); Air/Gas in Device (4062)
Patient Problems Abdominal Pain (1685); Nausea (1970); Vomiting (2144)
Event Date 03/08/2024
Event Type  malfunction  
Event Description
Hyperinflation of balloon with impact on the antrum, nausea, vomiting and abdominal distension.
 
Manufacturer Narrative
To date, spatz fgia inc.Has not received the product for evaluation, therefore no analysis or testing has been done.A review of the device labeling notes the following: spontaneous hyperinflation is the enlargement of the balloon with extra air that can occur spontaneously.This can lead to symptoms such as pain, nausea, vomiting, dehydration, ulceration, perforation, and could require a down adjustment or removal of the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPATZ3 ADJUSTABLE BALLOON SYSTEM
Type of Device
GASTRIC BALLOON
Manufacturer (Section D)
SPATZ FGIA INC.
1801 s perimeter rd, ste 130
fort lauderdale FL 33309
Manufacturer (Section G)
SPATZ FGIA INC.
1801 s perimeter rd, ste 130
fort lauderdale FL 33309
Manufacturer Contact
pnina polishuk
1801 s perimeter rd, ste 130
fort lauderdale, FL 33309
MDR Report Key18900959
MDR Text Key337697633
Report Number3012638928-2024-03055
Device Sequence Number1
Product Code LTI
UDI-Device Identifier860005178810
UDI-Public(01)860005178810(11)2023-03-16(17)2025-03-16(10)230316
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P190012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA-SP3-03K
Device Lot Number230316
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age36 YR
Patient Weight116 KG
-
-