MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER
|
Back to Search Results |
|
Model Number PED2-450-18 |
Device Problems
Retraction Problem (1536); Activation Failure (3270)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id fg15150-0615-1s (lot: 227122907); product type: ; implant date n/a; explant date n/a medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received a report that the pipeline failed to open and during retrieval got stuck in the catheter.The patient was undergoing blood flow directing dense mesh stent implantation for treatment of a saccular, unruptured aneurysm in c6 segment with a max diameter of 7.1 mm and a 5.2 mm neck diameter.Landing zone: distal: 4.0 mm and proximal: 4.5 mm.The accessed vessel was the femoral artery with a diameter of 6 mm.It was noted the patient's vessel tortuosity was moderate.The angiographic result post procedure showed smooth blood flow.It was reported that during aneurysm surgery, the tip of the stent came out in an eccentric umbrella shape with the back tightened, and the stent could not be opened after many attempts.When withdrawing the system, the stent got stuck in the microcatheter and could not be withdrawn.The microcatheter was damaged.Replace with another system, and the surgery ended successfully.The pipeline was not used for an indication that is off-label.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).The catheter was flushed as indicated in the ifu.No patient symptoms or further complications were reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information was received that the pipeline was placed in a straight section when it failed to open.There were no other instruments replaced, and the device still couldn't be opened after multiple attempts.
|
|