PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MP5 Back to Search Results
Model Number 865024
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Event Description
The customer reported when "alarm reset" is pressed, the next "alarm stop" response is made.The device was in use on a patient.There was no report of patient or user harm.
 
Manufacturer Narrative
The remote service engineer (rse) spoke to the customer and advised them to go into service mode and calibrate the touch panel, but that did not improve the situation.A field service engineer (fse) went onsite and determined that the touchscreen required replacement.The device was operational after replacing the touchscreen.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.Reporter phone number: (b)(6) reporting institution phone number: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MP5
Type of Device
INTELLIVUE MP5
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18902857
MDR Text Key337649771
Report Number9610816-2024-00143
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838000230
UDI-Public00884838000230
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865024
Device Catalogue Number865024
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/27/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-