THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; Control, pump speed, cardiopulmonary bypass Back to Search Results
Model Number 201-90411
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Manufacturer Narrative
A1-a4 patient information requested but not provided.No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
It was reported that while a centrimag system was being used on a patient, the customer observed an alarm.The centrimag pump did not stop and the patient was not impacted.The customer swapped out the centrimag console and centrimag motor with new ones.The customer also tested the console with new motors and it worked as expected.The customer tested the motor with new consoles and the motor alarmed on each of the consoles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of Device
Control, pump speed, cardiopulmonary bypass
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18904881
MDR Text Key337756456
Report Number3003306248-2024-00438
Device Sequence Number1
Product Code DWA
UDI-Device Identifier07640135140702
UDI-Public7640135140702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131179
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number201-90411
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-