• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; Pump, blood, cardiopulmonary bypass, non-roller type

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; Pump, blood, cardiopulmonary bypass, non-roller type Back to Search Results
Model Number 102956
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Event Description
It was reported that while a centrimag system was being used on a patient, the customer observed an alarm.The centrimag pump did not stop and the patient was not impacted.The customer swapped out the centrimag console and centrimag motor with new ones.The customer also tested the console with new motors and it worked as expected.The customer tested the motor with new consoles and the motor alarmed on each of the consoles.
 
Manufacturer Narrative
A1-a4 patient information requested but not provided.No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Type of Device
Pump, blood, cardiopulmonary bypass, non-roller type
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18904904
MDR Text Key337760731
Report Number3003306248-2024-00435
Device Sequence Number1
Product Code KFM
UDI-Device Identifier07640135140078
UDI-Public07640135140078
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102956
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-