Catalog Number 07P72-20
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a1 - patient identifier complete entry = (b)(6) all available patient information was included.Additional patient details are not available.
 
Event Description
The customer observed falsely elevated alinity c carbon dioxide results for five patients.The following data was provided (customer¿s reference range is 22-29 meq/l): sample id (b)(6) initial result was 34, repeat was 29 meq/l sample id (b)(6) initial result was 32, repeat was 26 meq/l sample id (b)(6) initial result was 34, repeat was 29 meq/l sample id (b)(6) initial result was 30, repeat results were 27 and 23 meq/l sample id (b)(6) initial result was 34, repeat was 26 meq/l.There was no impact to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C CARBON DIOXIDE REAGENT KIT
Type of Device
ENZYMATIC, CARBON-DIOXIDE
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18906916
MDR Text Key337697742
Report Number3002809144-2024-00074
Device Sequence Number1
Product Code KHS
UDI-Device Identifier00380740121594
UDI-Public00380740121594
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060295
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07P72-20
Device Lot Number65006UQ10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY C PROCESSING MODU, 03R67-01, AC04063
-
-