Catalog Number 10492730
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Manufacturer Narrative
Based on the information provided by the customer, the cause of this event was user error.The operators have been retrained the hospital has updated its' procedure to mandate that the patient case number be in the sticker barcode and that the barcode will be scanned at the beginning of the open-heart procedure and triple checked to the patient chart.No device problem.
 
Event Description
The customer reported that the incorrect patient id was selected 13 times during an open-heart procedure.The operator incorrectly attempted to use the "previous patient" feature.The results were sent to the incorrect patient chart.The error was identified and the results were edited and transferred to the correct patient chart.There is no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD.
northern road
chilton industrial estate
sudbury, suffolk CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
MDR Report Key18907095
MDR Text Key337758219
Report Number3002637618-2024-00017
Device Sequence Number1
Product Code CHL
UDI-Device Identifier00630414589169
UDI-Public00630414589169
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10492730
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/16/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-