MEDOS INTERNATIONAL SARL 5.5 HEALIX ADVANCE KNTLS PEEK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Catalog Number 222585 |
Device Problems
Break (1069); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/07/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.H4: the device manufacture date was unknown.Udi: (b)(4).
|
|
Event Description
|
It was reported by the sales rep that during a shoulder rotator cuff repair surgical procedure on (b)(6) 2024 the 5.5 healix advance kntls peek device anchor was broken, while try to tighten the suture and insert the anchor.Device is not available for evaluation.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4, h4: the expiration date and device manufacture date have been added.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.B5; during further follow up it was reported that the procedure was completed successfully.A new anchor was used and there was no delay to the procedure or patient harm.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the complaint device is not being returned, it was discarded, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information or the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device, and no non-conformance was identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|