• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 101967-950
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.The rma was authorized but not received, thus no confirmation or investigation of the complaint was possible.As part of resolution, the user received a new sensor and a transmitter.No further resolution was found necessary for this complaint.
 
Event Description
On (b)(6)2023,senseonics was made aware of an incident where user cannot see the glucose which led to sensor removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key18909706
MDR Text Key337739363
Report Number3009862700-2024-00539
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491023308
UDI-Public817491023308
Combination Product (y/n)Y
Reporter Country CodeIT
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/23/2023
Device Model Number101967-950
Device Catalogue NumberFG-4500-50-302
Device Lot NumberWP08965
Was Device Available for Evaluation? No
Date Manufacturer Received05/02/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-