ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 14687-15 |
Device Problems
Crack (1135); Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch cracked and generated a leak of an unspecified medication during patient use.The reporter stated, "plum set cracked which resulted in drug leakage." there was no drug exposure to the patient or healthcare provider.There was no healthcare provider harm.There was no patient harm reported.
|
|
Manufacturer Narrative
|
Received one (1) used list #14687-15 primary plum set.The set was attached to an icu medical provided intravenous (iv) bag and primed per packaging directions.A leak was noted to be coming from the cassette and no further infusion was attempted.The set was also air pressure leak tested and there were two leaks observed along the edge of the cassette.Additionally, the cassette underwent stack height measurements of which two corners failed.The complaint of plum set cracked which resulted in leakage can be confirmed.The probable cause of this event is related to the product being exposed to excessive heat during transportation, which may contribute to the warpage of the cassette leading to cassette errors and leakage.The lot history was reviewed, and no nonconformities were identified that may have contributed to the reported complaint.
|
|