• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 14687-15
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2024
Event Type  malfunction  
Event Description
It was reported that a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch cracked and generated a leak of an unspecified medication during patient use.The reporter stated, "plum set cracked which resulted in drug leakage." there was no drug exposure to the patient or healthcare provider.There was no healthcare provider harm.There was no patient harm reported.
 
Manufacturer Narrative
Received one (1) used list #14687-15 primary plum set.The set was attached to an icu medical provided intravenous (iv) bag and primed per packaging directions.A leak was noted to be coming from the cassette and no further infusion was attempted.The set was also air pressure leak tested and there were two leaks observed along the edge of the cassette.Additionally, the cassette underwent stack height measurements of which two corners failed.The complaint of plum set cracked which resulted in leakage can be confirmed.The probable cause of this event is related to the product being exposed to excessive heat during transportation, which may contribute to the warpage of the cassette leading to cassette errors and leakage.The lot history was reviewed, and no nonconformities were identified that may have contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18909736
MDR Text Key337766687
Report Number9615050-2024-00235
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10887787005650
UDI-Public(01)10887787005650(17)260801(10)13670553
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number14687-15
Device Lot Number13670553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNSPECIFIED FLUID, UNK MFR
-
-