CONMED LARGO HEX FLEX CANNULA WITH DISPOSABLE OBTURATOR, 8.0 X 85MM; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number C7480
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Event Description
The sales representative reported on behalf of the customer that the c7480, hex flex cannula with disposable obturator, 8.0 x 85mm, was being used during an unknown procedure on (b)(6) 2024 when it was reported, ¿the tip of the cannula breaks, causing fragments of the product to fall into the patient's joint cavity.It took a long time for the surgeon to remove the fragment.".There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event Description
The sales representative reported on behalf of the customer that the c7480, hex flex cannula with disposable obturator, 8.0 x 85mm, was being used during an unknown procedure on 27feb24 when it was reported, ¿the tip of the cannula breaks, causing fragments of the product to fall into the patient's joint cavity.It took a long time for the surgeon to remove the fragment.".There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Manufacturer Narrative
The device has not been returned for evaluation to date however, a photograph was provided.Review of the photograph however, does not show the tip of the cannula.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A 2 year lot history review shows a total of 2 devices for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 2 complaints, regarding 3 devices, for this device family and failure mode.During this same time frame 56,656 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.00005.Per the instructions for use, the user is advised the following: 1.Remove items from the sterile package using aseptic techniques.2.Place the obturator into the hex flex cannula.If using the reusable obturator, place the hex flex assembly over a switching stick for entry into the joint.Reusable obturator sold separately.3.Verify that the hex key on the obturator is engaged with the cap.4.Slowly screw the cannula clockwise when inserting it into the joint.Once in the correct position the cannula will engage the capsule, muscle, skin, etc.This will prevent the cannula from backing out of the joint.5.To remove the cannula from the joint, slowly screw the cannula counterclockwise.6.Properly discard the hex flex cannula after use per facility guidelines.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEX FLEX CANNULA WITH DISPOSABLE OBTURATOR, 8.0 X 85MM
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
CONMED LARGO
11311 concept blvd
largo FL 33773
Manufacturer (Section G)
CONMED LARGO
11311 concept blvd
largo FL 33773
Manufacturer Contact
john berga
11311 concept blvd
largo, FL 33773
7273995358
MDR Report Key18911338
MDR Text Key337744231
Report Number1017294-2024-00029
Device Sequence Number1
Product Code NBH
UDI-Device Identifier10845854018408
UDI-Public(01)10845854018408(17)280828(10)1330160
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC7480
Device Lot Number1330160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/30/2023
Is the Device Single Use? Yes
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-