Catalog Number 199725640S
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problems Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d4: lot # is either 219213, 224775, or 328933.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: initial reporter is a synthes employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from japan reports an event as follows: it was reported that on (b)(6) 2022, a posterior lumbar interbody fusion (l3/4) was performed.The procedure was completed successfully with no surgical delay.On (b)(6) 2024 a revision procedure was performed for an unknown cause.L3/4 plif had bony fusion.During the surgery, the sfx unitized set screw and rod were removed.When the l3 left screw in question was removed, it was found to have broken just below the screw head.The l3 right side screw was then removed, a cage was installed between l4/5, a screw was inserted at l5, and a rod, unitized set screw and sfx were installed.The distal screws from the screw head of the broken screw were retained in the body.There are no broken pieces in the patient, which was confirmed by x-ray.The surgery was completed successfully with no surgical delay.The patient outcome is stable.No further information is available.This report is for a 5.5 exp verse can scr 6.0x40.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE CAN SCR 6.0X40
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key18912650
MDR Text Key337765564
Report Number1526439-2024-01057
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034440668
UDI-Public(01)10705034440668
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K142185
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number199725640S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-