ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625374160
Device Problem Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2024
Event Type  malfunction  
Event Description
We received an allegation of questionable inr results with coaguchek xs meter serial number (b)(6).At 1:11 pm, the meter result was 3.7 inr.At 1:21 pm, the meter result was 2.6 inr.Customer's therapeutic range is 2.0-3.0 inr.The customer tests on a weekly basis.
 
Manufacturer Narrative
The meter and test strips were requested for return.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18913091
MDR Text Key337769250
Report Number1823260-2024-00771
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702127104
UDI-Public00365702127104
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number04625374160
Device Lot Number70302321
Was Device Available for Evaluation? No
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DILTIAZEM; TRAMADOL; UNSPECIFIED HEART PILL; WARFARIN
Patient Age79 YR
Patient SexMale
Patient Weight86 KG
-
-