• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FERRITIN; FERRITIN IMMUNOLOGICAL TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FERRITIN; FERRITIN IMMUNOLOGICAL TEST SYSTEM Back to Search Results
Catalog Number 07027273190
Device Problem Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Manufacturer Narrative
The cobas e 801 module serial number was (b)(6).The investigation is ongoing.
 
Event Description
There was an allegation of questionable ferritin elecsys e2g results from the cobas e 801 module.The result was 940 ng/ml.Using a different sample from the same venipuncture, the result from an abbott analyzer was 594 ng/ml.Using another cobas analyzer, the result was 866 ng/ml.
 
Manufacturer Narrative
The investigation did not identify a product problem.The specific cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FERRITIN
Type of Device
FERRITIN IMMUNOLOGICAL TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18914588
MDR Text Key337791089
Report Number1823260-2024-00774
Device Sequence Number1
Product Code JMG
UDI-Device Identifier04015630939688
UDI-Public04015630939688
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
K971833
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07027273190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-