Model Number 71992-01 |
Device Problems
Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Hemorrhage/Bleeding (1888); Device Embedded In Tissue or Plaque (3165); Skin Infection (4544)
|
Event Date 02/24/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer reported that after showing, it was observed that the abbott diabetes care (adc) device was loose and the "sensor tip was half out".The customer was assisted by a non-healthcare professional in removing the sensor however, part of the sensor tip was left in the skin and they experienced bleeding.The customer was seen at a hospital where a healthcare professional removed the remaining sensor tip, applied a bandage, and provided "threlone" (antibiotic) ointment for treatment.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Event Description
|
A customer reported that after showing, it was observed that the abbott diabetes care (adc) device was loose and the "sensor tip was half out".The customer was assisted by a non-healthcare professional in removing the sensor however, part of the sensor tip was left in the skin and they experienced bleeding.The customer was seen at a hospital where a healthcare professional removed the remaining sensor tip, applied a bandage, and provided "threlone" (antibiotic) ointment for treatment.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this incident are the adhesive, including the adhesive irritating the user¿s skin, or misuse, including improper site selection and repeatedly using the same application site to place the sensor.These conditions are mitigated through the freestyle product labelling.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned for this complaint.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.Dhrs (device history record) for the freestyle libre sensor and freestyle libre sensor kit were reviewed, and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned for this complaint.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.Dhrs (device history record) for the freestyle libre sensor and freestyle libre sensor kit were reviewed, and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.This serves as a correction report.Section h10 (addtl mfg narrative) was incorrectly documented in the previous report.Correction has been made.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer reported that after showing, it was observed that the abbott diabetes care (adc) device was loose and the "sensor tip was half out".The customer was assisted by a non-healthcare professional in removing the sensor however, part of the sensor tip was left in the skin and they experienced bleeding.The customer was seen at a hospital where a healthcare professional removed the remaining sensor tip, applied a bandage, and provided "threlone" (antibiotic) ointment for treatment.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|