MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS); Ventricular (assist) bypass

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS); Ventricular (assist) bypass Back to Search Results
Model Number 106015
Device Problems Mechanical Problem (1384); Decreased Pump Speed (1500); Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Manufacturer Narrative
The short to shield is documented under mfr - 2916596-2019-00591.Previous driveline revision is documented under mfr- 2916596-2023-00450.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Event Description
It was reported that the patient had a pump stop and it took an extended period of time (over 10 minutes) for the pump speed return to normal.It was noted that the patient had a known phase to phase internal short and was using an ungrounded cable.Log files were submitted for review and captured a pump stop on (b)(6) 2024.Log files were submitted for review showed 3 low speed advisory(s) and 2 pump stop events on (b)(6) 2024 at 8:46 am.Speed drops were as low as 0 revolutions per minute (rpm).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS)
Type of Device
Ventricular (assist) bypass
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18915209
MDR Text Key337848714
Report Number2916596-2024-01573
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024011224
UDI-Public00813024011224
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P060040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number106015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
Patient Weight66 KG
-
-