Catalog Number 11972103160 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The pth reagent expiration date was not provided.The cobas e411 rack serial number was (b)(6).The field service engineer found that the tip buffer sensor was dirty and the measuring cell was degraded.He cleaned the sensor and replaced the measuring cell.He then ran the assay testing and performed all checks and they were successful.The customer performed calibration and qc and they were acceptable.The investigation is ongoing.
|
|
Event Description
|
We received an allegation about discrepant results for 1 patient's plasma sample tested with elecsys parathyroid hormone (pth) assay on a cobas e411 immunoassay analyzer.Initial result: 10.17 pg/ml.Repeat result: 342.7 pg/ml.The initial result was automatically reported outside the laboratory.The technician then repeated the sample as the result did not match the patient's medical history.The technician reportedly called the doctor with the correct repeat result.
|
|
Manufacturer Narrative
|
The cause was the dirty tip buffer sensor in the second tip position and the degrading measuring cell.The service actions (cleaning the sensor and replacing the measuring cell) resolved the issue.No further issues were reported afterward.
|
|
Search Alerts/Recalls
|