• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC DAVINCI XI; SURGEON SIDE CONSOLE, SMART PEDALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC DAVINCI XI; SURGEON SIDE CONSOLE, SMART PEDALS Back to Search Results
Model Number 380677-06
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The isi fse was unable to reproduce the reported issue and replaced the master tool manipulator (mtm) as a precaution to resolve the issue.The system was tested and verified as ready for use.Isi has received the mtm; however, failure analysis has not completed their investigation.A follow-up mdr will be submitted when failure analysis has completed their investigation or if additional information is received.
 
Event Description
It was reported that during a da vinci-assisted paraesophageal hiatal hernia surgical procedure, the right master tool manipulators (mtm) were not clutching.Prior to calling in the issue, the customer elected to bring in a surgeon side console (ssc) from another system to complete the procedure.No troubleshooting was performed before converting to another dv system.The intuitive surgical, inc.(isi) technical support engineer (tse) the viewed system logs and noted error 23015 for the right mtm control.No known impact or patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVINCI XI
Type of Device
SURGEON SIDE CONSOLE, SMART PEDALS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18918311
MDR Text Key337848823
Report Number2955842-2024-12422
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874110744
UDI-Public(01)00886874110744
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number380677-06
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/04/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-