BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U128 |
Device Problems
Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Incorrect, Inadequate or Imprecise Result or Readings (1535); Inappropriate or Unexpected Reset (2959)
|
Patient Problems
Device Overstimulation of Tissue (1991); Pain (1994); Presyncope (4410); Asystole (4442)
|
Event Date 03/01/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the patient implanted with this cardiac resynchronization therapy pacemaker (crt-p) presented to the emergency room (er) due to pain and phrenic nerve stimulation.Noise with oversensing was observed with upper body movements, resulting in pacing inhibition with approximately 5-second intervals of asystole.The patient was nearsyncopal, but denied loss of consciousness.Device interrogation revealed that this crt-p had entered safety mode, and the patient was flown to a different facility for emergent device replacement.Device replacement was completed, and this crt-p was expected to be returned for analysis.No additional adverse patient effects were reported.
|
|