It was reported that the patient implanted with this implantable cardioverter defibrillator (icd) typically had atrial pacing with ventricular sensing, or ap vs.But whenever a premature ventricular contraction (pvc) occurred, the device started pacing and av conduction was not re-established.Programming assistance was requested, and the field representative noted that avsearchhysteresis (avsh) was off at the time.Review of a-a and v-v timing showed conduction was approximately 240 ms, and once a pvc occurred the device began av pacing.Avsh did not appear to be programmed off, as the first av interval was longer than the preceding av interval.The subsequent ap was likely delivered into refractory tissue, thus did not capture.Additionally, the patient blocked at rates above 100.Technical services (ts) recommended decreasing the lower rate limit (lrl) to 60 beats per minute (bpm) from 70 bpm and decreasing the avsh interval to 300-330 ms from 400 ms.There was also discussion about enabling atrial flutter response (afr).This icd remains in service.No adverse patient effects were reported.
|