Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The horizon clip applier instrument and failure analysis investigations not replicated nor confirmed the customer reported complaint.Failure analysis found the primary failure of could not reproduce to be related to the customer reported complaint.The instrument was placed and driven on an in-house system.The instrument passed the recognition and engagement tests.There was no report of recognition or engagement failure during procedure on the system as the instrument was not used and has 100 out of 100 lives in it.Additional observations not reported by site: the horizon clip applier instrument was found to have corrosion/contamination on the bearings.An input disk bearings exhibited orange discoloration.The instrument was found to have a broken distal pitch cable at the proximal clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.The cable was fully broken.There were evidence of discoloration or corrosion/ contamination on the cable to indicate a reprocessing induced issue.The components surrounding to the broken cable did not exhibit abnormal sharp edges or damage that would have contributed to the cable breaking.The instrument was found to have one of the pitch cables broken at the proximal end in the housing.The pitch cable was broken at the backend pulleys.The clamping pulley did exhibit signs of corrosion/contamination/ residue that would have contributed to the broken cable.The instrument was found to have an excessive amount of dried, white residue on the clamping pulley of inputs #[5 (pitch), 6 (grip) , 7 (grip) located in the backend of the instrument.The groove surfaces exhibited a residual, powder-like deposit, that could be removed by wiping.
|