Model Number G151 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/28/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this device record a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data was sent for engineering analysis.Analysis showed that the low voltage fault is valid and confirmed that the power consumption is increasing in a gradual fashion.Device replacement was suggested.To date, this device remains in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that this device record a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data was sent for engineering analysis.Analysis showed that the low voltage fault is valid and confirmed that the power consumption is increasing in a gradual fashion.Device replacement was suggested.To date, this device remains in service.No adverse patient effects were reported.Additional information received indicates that the device was explanted.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
|
|