• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING Back to Search Results
Model Number 865350
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information regarding the reported event and the investigation is ongoing.A follow-up report will be submitted upon completion of the investigation.
 
Event Description
The customer reported a speaker malfunction, no sound was coming from the device.The device was use at time of event, there was no adverse event reported.
 
Manufacturer Narrative
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that the speaker produced sound.Although the speaker was confirmed to be functioning per specification during testing it was indicated that there was no sound at the time of the event, the speaker has been replaced per current process.The device was operational after repairs were completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18921830
MDR Text Key337875158
Report Number1218950-2024-00195
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082236
UDI-Public00884838082236
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-