WILSON-COOK MEDICAL INC FUSION EXTRACTION BALLOON WITH MULTIPLE SIZING; GCA, CATHETER, BILIARY, SURGICAL
|
Back to Search Results |
|
Catalog Number FS-8.5-12-15-A |
Device Problem
Deflation Problem (1149)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2024 |
Event Type
malfunction
|
Event Description
|
In preparation for a procedure, the user selected a cook fusion extraction balloon with multiple sizing.It was reported [that] the extraction balloon inflates but it remains blocked; it does not deflate.This occurred prior to patient contact; there was no impact to the patient.
|
|
Manufacturer Narrative
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all fusion extraction balloon with multiple sizing are subjected to a visual and functional test to ensure device integrity.The functional test includes an air inflation test to ensure proper balloon function.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
Manufacturer Narrative
|
Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved could not confirm the report as it was described.The balloon inflated and deflated without difficulty.The device was returned without the syringe and stopcock.A functional test was performed on the balloon using a syringe and stopcock from our stock; when the balloon was filled with air, no leaks were present in the balloon material.Once the balloon was inflated, pressure was released from the syringe and the balloon deflated within a few seconds (reference attached inflation/deflation video).The device was visually inspected and appeared to be free of bends and kinks along the catheter of the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the device could not confirm the complaint, the balloon inflated and deflated without difficulty.A definitive cause for this observation could not be determined because the actual prior to use testing conditions could not be duplicated in the laboratory setting as the syringe and stopcock were not returned.This limits our ability to conclusively determine a cause.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.Prior to distribution, all fusion extraction balloon with multiple sizing are subjected to a visual and functional test to ensure device integrity.The functional test includes an air inflation test to ensure proper balloon function.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
Search Alerts/Recalls
|
|
|