It was reported that the right ventricular (rv) lead exhibited high out of range pacing impedance measurements of greater than 2000 ohms along with increasing threshold measurements.The patient underwent provocation maneuvers and no oversensing or noise was observed.It was believed that the change in values is due to the age of the lead as it has been implanted for 24 years.The output was increased to maximum value.Data was sent into technical services (ts) for review.Upon review ts noted that the increase in impedance measurement has been gradual.Ts discussed higher thresholds need to be evaluated for appropriate programming of outputs to ensure appropriate capture.Ts indicated that this could be a symptom of encapsulation or ionization, but further testing would need to occur.No noise episodes have been stored, however ts observed greater than three thousand counters that were greater than 250 minutes during a six-month time frame.Ts discussed troubleshooting the lead to determine the reason for the high counters.The rv lead remains in service and no adverse effects were reported.
|